当前位置:上海境宣实业有限公司>>日本光明北川式气体检测管>>气体快速检测管>> 134SB三氯乙烯Cl2C=CHCl气体测毒管134SB TRICHLOROETHYLENE 0.2-36.8
三氯乙烯Cl2C=CHCl气体测毒管134SB TRICHLOROETHYLENE 0.2-36.8ppm
1. PERFORMANCE
1) Measuring range :1-16 ppm 0.2-3.2 ppm
Number of pump strokes 1(100ml) 4(400ml)
2) Sampling time :1.5 minutes/1 pump stroke
3) Detectable limit :0.1 ppm(400ml)
4) Shelf life :1 year (Necessary to store in a refrigerated place ; 0~10℃)
5) Operating temperature :0~40℃
6) Reading :Direct reading from the scale calibrated by 1 pump stroke
7) Colour change :Pale orange→Blueish purple
2. RELATIVE STANDARD DEVIATION
RSD-low:10% RSD-mid.:10% RSD-high:10%
3. CHEMICAL REACTION
By reacting with an Oxidizer, Hydrogen chloride is produced and PH indicator is discoloured.
CI2C=CHCI+PbO2+H2SO4→HCI
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
GAS
To pump
OXIDIZER
(White)
16
12
14
10
8
6
4
2 1
1000ml
ppm
Tube No.
134SB TRICHLOROETHYLENE
Substance Interference ppm Coexistence
Tetrachloroethylene FIG.1 Similar stain is produced. 2 Higher readings are given.
1,2-Dichloroethylene FIG.2 〃2 〃
Hydrogen chloride 〃2 〃
Vinyl chloride 〃40 〃
0
15
Tetrachloroethylene〔ppm〕�
FIG.1 Influence of Tetrachloroethulene
Trichloroethylene indicating value 〔ppm〕
10
10
5
5
Coexistence
Interference
0
15
1,2-Dichloroethylene〔ppm〕�
FIG.2 Influence of 1,2-Dichloroethylene
Trichloroethylene indicating value 〔ppm〕
40
10
20
5
Coexistence
Interference
(NOTE)
In case of 4 pump strokes, following formula is available for the actual concentration.
Actual concentration=1/5×Reading value.