三氯乙烯Cl2C=CHCl气体检测管134SA TRICHLOROETHYLENE 5-300ppm
1. PERFORMANCE
1) Measuring range :10-300 ppm 5-150 ppm
Number of pump strokes 1/2(50ml) 1(100ml)
2) Sampling time :2 minutes/1 pump stroke
3) Detectable limit :1 ppm(100ml)
4) Shelf life :2 years(Necessary to store in a refrigerated place ; 0~10℃)
5) Operating temperature :0~40℃
6) Temperature compensation :Necessary(See "TEMPERATURE CORRECTION TABLE")
7) Reading :Direct reading from the scale calibrated by 1 pump stroke
8) Colour change :Yellow→Red
2. RELATIVE STANDARD DEVIATION
RSD-low:10% RSD-mid.:10% RSD-high:10%
3. CHEMICAL REACTI0N
By decomposing with an Oxidizer, Hydrogen chloride is produced and PH indicator is discoloured.
CI2C=CHCI+PbO2+H2SO4→HCI
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
GAS To pump
150
100
50
30
20
10
5
ppm
PRETREAT REAGENT
(Grey)
Tube No.
134SA TRICHLOROETHYLENE
(NOTE)
In case of 1 / 2 pump strokes, following formula is available for the actual concentration.
Actual concentration=2×Temperature corrected value
Substance Interference Coexistence
Vinyl chloride FIG.2 Similar stain is produced. Higher readings are given.
Hydrogen chloride 〃〃
1,2-Dichloroethylene 〃〃
Tetrachloroethylene 〃〃
Chlorine FIG.1 Pale red stain is produced.
0
200
100
Chlorine〔ppm〕�
FIG.1 Influence of Chlorine
Trichloroethylene indicating value 〔ppm〕
50 100
Coexistence
Interference
0
200
100
Vinyl chloride〔ppm〕�
FIG.2 Influence of Vinyl chloride
Trichloroethylene indicating value 〔ppm〕
200 400
Coexistence
Interference
TEMPERATURE CORRECTION TABLE
Scale True Concentration (ppm)
Readings 0℃ 10℃ 20℃ 30℃ 40℃
(ppm) (32°F) (50°F) (68°F) (86°F) (104°F)
150 - 162 150 144 142
100 120 108 100 96 94
50 58 53 50 48 46
30 34 32 30 29 28
20 20 20 20 20 20