当前位置:上海境宣实业有限公司>>日本光明北川式气体检测管>>气体快速检测管>> 119SA甲醇CH3OH气体检测管119SA METHYL ALCOHOL(METHANOL) 0.05-6%
甲醇CH3OH气体检测管119SA METHYL ALCOHOL(METHANOL) 0.05-6.0%
Tube 19SA
1. PERFORMANCE
1) Measuring range :0.05-6.0%
Number of pump strokes 1(100ml)
2) Sampling time :1.5 minutes/1 pump stroke
3) Detectable limit :100 ppm
4) Shelf life :3 years
5) Operating temperature :0~40℃
6) Temperature compensation :Necessary(See "TEMPERATURE CORRECTION TABLE")
7) Reading :Direct reading from the scale calibrated by 1 pump stroke
8) Colour change :Yellow orange→Pale green
2. RELATIVE STANDARD DEVIATION
RSD-low:10% RSD-mid.:5% RSD-high:5%
3. CHEMICAL REACTION
Potassium dichromate is reduced
CH3OH+Cr6++H2SO4→Cr3+
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
0.05
0.1
0.5
1.0
2.0
3.0
4.0
5.0
6.0
%
GAS To pump
DEHUMIDIFIER
(White)
119SA
METHYL ALCOHOL
(METHANOL)
Substance Interference ppm Coexistence
Paraffin (morefi nth haynd Cro3c)arbons FIG.1 Similar stain is produced. Higher readings are given.
Alcohols FIG.2 〃〃
Esters 〃50 〃
Ketones 〃〃
Aromatic hydrocarbons 〃〃
Reagent is discoloured to If the top of the discolouration can
Halogenated hydrocarbons Brown slightly. ibneg so bist aninoet da,fftehcet eadc.curacy of read-
(NOTE)
Although the top of light green stain changes to Brown gradually, read the concentration on the top of the light
green stain shortly.
0
5
6
3
4
Hexane〔%〕�
FIG.1 Influence of Hexane
Methanol indicating value 〔%〕
0.1 0.2 0.3
2
1
0
5
6
3
4
Ethanol〔%〕�
FIG.2 Influence of Ethanol
Methanol indicating value 〔%〕
0.1 0.2 0.3 0.4 0.5
2
1
Coexistence
Interference
TEMPERATURE CORRECTION TABLE
Scale True Concentration (%)
Readings 0℃ 10℃ 20℃ 30℃ 40℃
(%) (32°F) (50°F) (68°F) (86°F) (104°F)
6.0 6.0 4.3 3.3
5.0 5.0 3.7 2.8
4.0 4.0 3.0 2.3
3.0 5.1 3.0 2.3 1.8
2.0 3.0 2.0 1.6 1.2
1.0 2.7 1.4 1.0 0.8 0.6
0.5 1.0 0.7 0.5 0.4 0.3
0.1 0.1 0.1 0.1 0.1 0.1
请输入账号
请输入密码
请输验证码
以上信息由企业自行提供,信息内容的真实性、准确性和合法性由相关企业负责,化工仪器网对此不承担任何保证责任。