Troubleshooting Giude

【详细说明】

Troubleshooting Giude

Pipet-Aid Model:ProblemSolution
4-000-110 & 220No aspiration 1. Check valve placed in upside down, black o-ring goes up into the handle.
No aspiration1. Check valve in nosepiece needs to be cleaned out. Push wire or paperclip through black o-ring to open up check valve. When opened, rinse float and shell with water or alcohol. Allow to dry and reassemble.
2. Shake and listen for float to rattle. Reassemble into nosepiece and check for aspiration.
3. Tubing leading to vacuum side of pump has a hole in it. Cut out hole or replace tuging (4-000-004)
4. Hose connector missing or split. Replace hose connector (4-000-011)
Pipet leaks after filling1. Replace rubber insert (4-000-002)
2. Check tubing for holes, replace or shorten if found. Hose connector missing or split.
3. Replace hose connector(4-000-011)
4. Diaphragm inside pump is split or dried up and cracked, Replace diaphragms (4-000-010).
Fill/empty very slow but no leaking1. Rubber insert has cotton plugs blocking flow, clean out cotton plugs or replace rubber insert (4-000-002)
2. Clean out check valve or replace (4-000-003)
3. Buttons in valve body dirty, clean or replace.
4. Diaphragm inside pump is split or dried up and cracked, Replace diaphragms (4-000-010).
 
4-000-110-TC
& 220-TC
No aspiration1. Check valve placed in upside down, black o-ring goes up into the handle.
No aspiration1. Filter has been wetted. Replace with 4-000-051
2. TC gasket missing in nosepiece, requires one on each end of filter (4-000-057).
3. Tubing leading to vacuum side of pump has a hole in it. Cut out hole or replace tuging (4-000-004)
Pipet leaks after filling1. Replace rubber insert (4-000-002)
2. Check tubing for holes, replace or shorten if found.
3. Hose connector missing or split. Replace hose connector (4-000-011).
4. Diaphragm inside pump is split or dried up and cracked, Replace diaphragms (4-000-010).
Aspiration/fill very slow but no leaking1. Rubber insert has cotton plugs blocking flow, clean out cotton plugs or replace rubber insert (4-000-002)
2. Buttons in valve body dirty, clean or replace.
3. Diaphragm inside pump is split or dried up and cracked, Replace diaphragms (4-000-010).
4. Poor seal around filter; use Drummond filters (4-000-051).
 
4-000-111 & 211No aspiration1. Check valve placed in upside down, black o-ring goes up into the handle.
No aspiration1. Check valve in nosepiece needs to be cleaned out. Push wire or paperclip through black o-ring to open up check valve. When opened, rinse float and shell with water or alcohol. Allow to dry and reassemble.
2. Shake and listen for float to rattle. Reassemble into nosepiece and check for aspiration.
3. Tubing leading to vacuum side of pump has a hole in it. Cut out hole or replace tuging (4-000-004)
4. Hose connector missing or split. Replace hose connector (4-000-011)
 
Pipet leaks after filling1. Replace rubber insert (4-000-002)
2. Check tubing for holes, replace or shorten if found
3. Hose connector missing or split. Replace hose connector (4-000-011)
4. Diaphragm inside pump is split or dried up and cracked, Replace diaphragms (4-000-010).
Aspiration/fill very slow but no leaking1. Rubber insert has cotton plugs blocking flow, clean out cotton plugs or replace rubber insert (4-000-002)
2. Clean out check valve or replace (4-000-003)
3. Buttons in valve body dirty, clean or replace.
4. Diaphragm inside pump is split or dried up and cracked, Replace diaphragms (4-000-010).
 
4-000-111-TC
& 211-TC
No aspiration1. Check valve placed in upside down, black o-ring goes up into the handle.
No aspiration1. Filter has been wetted. Replace with 4-000-051
2. TC gasket missing in nosepiece, requires on on each end of filter (4-000-057).
3. Tubing leading to vacuum side of pump has a hole in it. Cut out hole or replace tuging (4-000-004)
Pipet leaks after filling1. Replace rubber insert (4-000-002)
2. Check tubing for holes, replace or shorten if found.
3. Hose connector missing or split. Replace hose connector (4-000-011)
4. Diaphragm inside pump is split or dried up and cracked, Replace diaphragms (4-000-010).
Aspiration/fill very slow but no leaking1. Rubber insert has cotton plugs blocking flow, clean out cotton plugs or replace rubber insert (4-000-002)
2. Buttons in valve body dirty, clean or replace.
3. Poor seal around filter; use Drummond filters (4-000-051)
4. Diaphragm inside pump is split or dried up and cracked, Replace diaphragms (4-000-010).
5. Poor seal around filter; use Drummond filters (4-000-051)
 
4-000-100 & 200No aspiration1. Filter has been wetted. Replace with 4-000-051
2. TC gasket missing in nosepiece, requires on on each end of filter (4-000-057).
No sound from motor when buttons depressed1. Battery needs to be charged overnight
2. Battery dead, replace with (4-000-035)
3. Listen for slight click when buttons are first pushed, if not heard, microswitch is broken, replace wire harness (4-000-046).
Aspiration/fill very slow but no leaking1. Rubber insert has cotton plugs blocking flow, clean out cotton plugs or replace rubber insert (4-000-002)
2. Buttons in valve body dirty, clean or replace.
Aspiration slow1. Poor seal around filter; use Drummond filters (4-000-051)
2. Rubber insert has cotton plugs blocking flow, clean out cotton plugs or replace rubber insert (4-000-002)
Unit leaks1. Replace rubber insert (4-000-002)
2. Poor seal around filter; use Drummond filters (4-000-051)
3. Check nosepiece for tight fit, if loose, replace (4-000-053-A)
Buttons damaged1. Replace buttons (4-000-019)
2. Handle connector tubing has a pin hole in it, cut out if long enough or replace (4-000-013)
Aspiration is very slow; dispensing may be slow or normal.Probable Cause: Liquid has been aspirated into the pump, which causes the flapper valve to stick and not function properly.

1. Open up the handle by removing the nosepiece and the three screws holding the handle halves together.
2. Take out the pump by disconnecting it from the tubing and the electrical connector.
3. Unscrew the four screws that hold down the two dark sections of the pump.
4. Remove the upper section from the next lower section. The upper section has the two ports extending from it. Once removed, this will expose a semi-transparent rubber flapper valve.
5. Gently remove this piece, clean it and the chamber with water. Dry both pieces & then replace the flapper valve.
6. This flapper valve can only be installed one way. Use the locating pins to align the valve. There is one large pin and one small pin to prevent misalignment and improper orientation of the valve.
7. Reassemble and reinstall. Check for aspiration speed.

To prevent this, the use of Drummond filters minimized the chance for damaging the valves but it may not always eliminate the problem.
 
4-000-101 & 201No aspiration1. Filter has been wetted. Replace with 4-000-051
2. TC gasket missing in nosepiece, requires on on each end of filter (4-000-057).
3. "Fill" button between speed settings, Rotate to you feel it click into place.
No sound from motor when buttons depressed1. Battery needs to be charged overnight
2. Battery dead, replace with (4-000-035)
3. Battery damaged by use of non-Drummond charger. Replace battery (4-000-035)
Aspiration/fill very slow but no leaking1. Rubber insert has cotton plugs blocking flow, clean out cotton plugs or replace rubber insert (4-000-002)
2. Buttons in valve body dirty, clean or replace.
Unit leaks1. Replace rubber insert (4-000-002)
2. Poor seal around filter; use Drummond filters (4-000-051)
3. Valve body o-rings need to be replaced. Call Drummond 1-
Aspiration is very slow; dispensing may be slow or normal.Probable Cause: Liquid has been aspirated into the pump, which causes the flapper valve to stick and not function properly.

1. Open up the handle by removing the nosepiece and the three screws holding the handle halves together.
2. Take out the pump by disconnecting it from the tubing and the electrical connector.
3. Unscrew the four screws that hold down the two dark sections of the pump.
4. Remove the upper section from the next lower section. The upper section has the two ports extending from it. Once removed, this will expose a semi-transparent rubber flapper valve.
5. Gently remove this piece, clean it and the chamber with water. Dry both pieces & then replace the flapper valve.
6. This flapper valve can only be installed one way. Use the locating pins to align the valve. There is one large pin and one small pin to prevent misalignment and improper orientation of the valve.
7. Reassemble and reinstall. Check for aspiration speed.

To prevent this, the use of Drummond filters minimized the chance for damaging the valves but it may not always eliminate the problem.
 
4-000-105No aspiration1. Filter has been wetted. Replace with 4-000-051
2. TC gasket missing in nosepiece, requires on on each end of filter (4-000-057).
3. "Fill" button between speed settings, Rotate to you feel it click into place.
No sound from motor when buttons depressed1. Plug in power supply/charger and use unit while it charges, use only power supply/charger 4-000-065. If battery will not hold a charger replace with new battery,(4-000-036) NOTE: Not the same battery as in the other portable units.
 
Aspiration/fill very slow but no leaking1. Rubber insert has cotton plugs blocking flow, clean out cotton plugs or replace rubber insert (4-000-002)
2. Buttons in valve body dirty, clean or replace.
Unit leaks1. Replace rubber insert (4-000-002)
2. Poor seal around filter; use Drummond filters (4-000-051
3. Valve body o-rings need to be replaced. Call Drummond 1-
 
4-000-300 & 302 & 303 No aspiration1. Filter has been wetted. Replace with 4-000-051
2. TC gasket missing in nosepiece, requires on on each end of filter (4-000-057).
3. "Fill" button between speed settings, Rotate to you feel it click into place.
Unit leaks1. Replace rubber insert (4-000-002)
2. Poor seal around filter; use Drummond filters (4-000-051)
3. Valve body o-rings need to be replaced. Call Drummond 1-
No sound when handle removed from cradle1. Actuator button is stuck in the down position.
Aspiration/fill very slow but no leaking1. Rubber insert has cotton plugs blocking flow, clean out cotton plugs or replace rubber insert (4-000-002)
2. Buttons in valve body dirty, clean or replace.
 
4-000-400 & 400-CUnit will not turn "on"1. Hold menu button down until it flashes and turns on.
2. Unit not plugged into the power supply
3. On portable unit using batteries, the battery is not charged.
Aliquot inaccurate1. Confirm you have the correct pipet settings (volume and brand)
2. Confirm you are using the correct filters, part number 4-000-051-E (essential for accuracy)
Programmed aliquot is inaccurate1. Perform CAL. Scroll to OPTNS, press menu; cursor will be on CAL, press menu; scroll to YES, press menu; remove any pipets, press menu.
Programmed aliquot is inaccurate but consistant1. Scroll to OPTNS, press menu; scroll to OFST, press menu; scroll to YES, press menu; if aliquot is too small, lower OFST value and check/if aliquot is too large, increase OFST value and check.

 

 

    
留言框
感兴趣的产品: *
留言内容:
您的姓名: *
您的单位:
联系电话: *
微信:
常用邮箱:
详细地址:
省份: *
验证码: * =   请输入计算结果(填写阿拉伯数字),如:三加四=7
      

产品搜索:

联系方式
地址:上海市虹口区花园路66号嘉和国际大厦1号楼26楼F座
邮编:200083
联系人:兰迪
电话:86-021-55221162/4009208916
传真:
手机:13661992907
留言:发送留言
网址:
商铺:https://www.chem17.com/st243535/
| 商铺首页 | 公司档案 | 产品展示 | 供应信息 | 公司动态 | 询价留言 | 联系我们 | 会员管理 |
化工仪器网 设计制作,未经允许翻录必究.Copyright(C) https://www.chem17.com, All rights reserved.
以上信息由企业自行提供,信息内容的真实性、准确性和合法性由相关企业负责,化工仪器网对此不承担任何保证责任。
温馨提示:为规避购买风险,建议您在购买产品前务必确认供应商资质及产品质量。
二维码 在线交流

扫一扫访问手机站