您好, 欢迎来到化工仪器网

| 注册| 产品展厅| 收藏该商铺

400-875-1717转801

Download

首页   >>   资料下载   >>   O determination in copper samples

弗尔德(上海)仪器设备有...

立即询价

您提交后,专属客服将第一时间为您服务

O determination in copper samples

阅读:5515      发布时间:2015-5-15
分享:
  Suitable analyzers
n ON 900
n OH 900
n ONH 2000
Used accessories
n Graphite crucibles (90190)
n Suitable calibration material (NIST or other)
Settings
n Comparator level: 20 mV
n Minimum time: 15 sec
n Maximum time: 1:30 min
n Post waiting: 10 sec
n Base line deviation: 200 mV;
Step: 16 mV; Time: 10 sec
n Mode: continuous
n Outgas
Time: 45 sec
Power: 3.5 kW
n Time
Purge: 20 sec
Stability: 45 sec
n Integration delay
IR cell: 2 sec
TC cell: 10 sec
n Analyse (1)
Time: 75 sec
Power (1): 3.0 kW
Power (2): 3.0 kW

Sample preparation
Make sure that the surface of the copper is free from
contaminations; otherwise clean the sample with acetone
p.a. and let dry at atmosphere. In case of very low oxygen
concentrations (< 5 ppm) rinse the sample in nitric acid
(HNO3) first, than water, last with acetone. Let this sample
dry at atmosphere and analyze immediay.
Procedure
n Prepare ELTRA analyzer (exchange anhydrone, sodium
hydroxide, copper oxide or Schuetze reagent when
necessary), clean furnace, sample drop mechanism,
electrode tip
n Run three blanks with empty crucibles
n Calibrate the analyzer with suitable calibration
material (NIST or other)
(1) Place empty crucible (90190) on the electrode tip,
close furnace (F2 Button)
(2) Weigh calibration material (usually pins)
(3) Place calibration material in the sample drop
mechanism and start analysis (F5 Button)
Repeat steps (1) – (3) at least three times;
Mark the results and use the calibration function in the
software.


提供商

弗尔德(上海)仪器设备有限公司

下载次数

269次

资料大小

343KB

资料类型

PDF 文件

资料图片

--

浏览次数

5515次

产品展示

会员登录

请输入账号

请输入密码

=

请输验证码

收藏该商铺

标签:
保存成功

(空格分隔,最多3个,单个标签最多10个字符)

常用:

提示

您的留言已提交成功!我们将在第一时间回复您~
在线留言